摘要: Double parallel-plate avalanche counters (DPPACs), each comprising two PPACs with delay-line readout for two-dimensional position determination, have been developed at Institute of Modern Physics, Chinese Academy of Sciences. The detector was tested with a standard α source, achieving a position resolution of about 0.4 mm. Each PPAC measures particle hit positions in two dimensions, with the dual configuration enhancing overall detection efficiency. Two DPPACs were deployed in a month-long experimental campaign at the Radioactive Ion Beam Line in Lanzhou (RIBLL1) with secondary beams of 10C, 10Be, 11C and 16C at typical intensities of 1.2 -- 1.5 × 105 Hz, yielding a stable single DPPAC-detection efficiency >90% and a typical in-beam position resolution ≈ 0.54 mm.
-
来自:
Yuan, Ms. Chen
-
分类:
物理学
>>
核物理学
-
备注:
已向《Nuclear Science and Techniques》投稿
-
引用:
ChinaXiv:202603.00098
(或此版本
ChinaXiv:202603.00098V1)
DOI:10.12074/202603.00098
CSTR:32003.36.ChinaXiv.202603.00098
-
科创链TXID:
8d19ff19-cb97-4b30-bd7d-fe71516be043
- 推荐引用方式:
Yuan, Ms. Chen,Shi, Mr. Fushuai,Xu, Dr. Xiaodong,Zhang, Dr. Zhichao,Liu, Dr. Enqiang,Yang, Dr. Herun,Ma, Dr. Peng,Duan, Prof. Limin 段利敏,ONG, Prof. HOOI-JIN ,Terashima, Dr. Saturo,Du, Mr. Ze-Yu,Hu, Mr. zhiheng,Huang, Mr. Taisen,LI, Mr. Lu,Li, Dr. PengJie,Liu, Dr. Hongna,Lv, Dr. Bingfeng,Ma, Mr. Junbing,Ma, Mr. Zixu,Sun, Dr. Yelei,Wang, Dr. H.,Wei, Dr. Xianglun,Yang, Prof. ZaiHong,Zhang, Mr. Weidong.Development and Application of Position-Sensitive Double Parallel-Plate Avalanche Counter.null.[DOI:10.12074/202603.00098]