Dosimetric Feasibility and Optimization of Boron-loaded Gel Lattices for Boron Neutron Capture Therapy

请选择邀稿期刊:
摘要: Purpose: To formulate a dose-design framework for boron-loaded gel lattices in boron neutron capture therapy and evaluate spatial dose structure, concentration allocation and irradiation direction in controlled analytic models.
Methods: The framework specifies concentration-direction-exposure selection, hybrid and discrete allocation, concentration sensitivity, beamlet control and spatial dose readback. Independent evaluation used three nine-insert analytic phantoms and a fixed 10-keV neutron source. Four equal-boron arms compared uniform single-field, concentration-selected single-field, uniform two-field and joint concentration-field selection. Response-based screening was followed by direct Monte Carlo evaluation of selected designs. A separate 2-mm baseline mesh supplied dose maps, a fixed-path peak-to-valley dose ratio and cumulative volume distributions. Additional methods were disclosed through source-linked formulations and synthetic software examples, separately from transport validation; no patient outcomes were included.
Results: The four-arm comparison yielded nine distinct physical plans across three phantoms. Direct calculations gave minimum-to-maximum insert-mean dose ratios of 0.411-0.472 for uniform single-field plans, 0.585-0.596 after single-field concentration redistribution, and 0.871-0.898 for joint selection. In every phantom, joint selection retained uniform concentration and reproduced the direction-only plan. At a matched minimum insert-mean dose, concentration redistribution reduced the normal-region mean dose point estimate in all three phantoms; joint selection increased it in two and decreased it in the third, relative to uniform single-field irradiation. The separate baseline path-defined peak-to-valley dose ratio was 4.96.
Conclusions: The calculations support a spatial lattice dose pattern and improved insert-dose balance under the stated model conditions, with geometry-dependent surrounding-tissue costs. Joint selection added no concentration benefit over the selected two-field plans. The broader disclosed methods require further validation; these results do not establish clinical efficacy or superiority to photon or proton therapy.

版本历史

[V1] 2026-11-21 03:37:36 ChinaXiv:202611.00215V1 下载全文
点击下载全文
预览
同行评议状态
待评议
许可声明
metrics指标
  •  点击量62
  •  下载量4
评论
分享
申请专家评阅
  • 运营单位: 中国科学院文献情报中心
  • 邮箱: eprint@mail.las.ac.cn
  • 地址:北京中关村北四环西路33号
招募志愿者 许可声明 法律声明

京ICP备05002861号-25 | 京公网安备11010802041489号
版权所有© 2016-2026 中国科学院文献情报中心