您当前的位置: > 详细浏览

The development of 222Rn detectors for JUNO prototype 后印本

请选择邀稿期刊:
摘要: The radioactive noble gas 222Rn, which can be dissolved in water, is an important background source for JUNO. In this paper, based on the water system of JUNO prototype, two kinds of high sensitivity radon detectors have been proposed and validated. The sensitivity of Si-PIN Rn detector, which uses Si-PIN photodiode to detect the -214Po, is 4.2 mBq/m3 and the sensitivity of LS Rn detector, which uses liquid scintillator to detect the coincident signals of -214Bi and -214Po, is 34.4 mBq/m3. Both of the two kinds of Rn detector have the potential to be developped into an online Rn concentration monitoring equipment for JUNO veto detector.

版本历史

[V1] 2017-11-10 15:12:52 ChinaXiv:201711.02258V1 下载全文
点击下载全文
预览
同行评议状态
待评议
许可声明
metrics指标
  •  点击量3404
  •  下载量1307
评论
分享
  • 运营单位: 中国科学院文献情报中心
  • 邮箱: eprint@mail.las.ac.cn
  • 地址:北京中关村北四环西路33号
招募志愿者 许可声明 法律声明

京ICP备05002861号-25 | 京公网安备11010802041489号
版权所有© 2016-2026 中国科学院文献情报中心