Monte Carlo calculation of PTW ionization chamber electrode correction factors in absorbed dose to water determination for 6 MV photon radiotherapy

请选择邀稿期刊:
摘要: Ionization chambers (ICs) are the primary reference instruments for measuring absorbed dose to water (D_w) in clinical photon beams from medical linear accelerators (Linacs). However, because IC electrodes are fabricated from non-water-equivalent materials, thus investigation of aggregated perturbation correction factors for the central and outer electrodes (f_c) of ICs is a crucial task before applying ICs in use. This work applied PHITS code to calculate f_c values for four Freiburg PTW ICs - i.e., PTW-31021 (0.070 cm³), PTW-31010 (0.125 cm³), PTW-31013 (0.300 cm³), and PTW-30010 (0.600 cm³). The beam spectrum was reconstructed based on the IAEA-distributed phase-space data corresponding to a 6 MV photon beam from an Elekta Precise Linac. Reference conditions followed established clinical protocols (IAEA-TRS 398): 10 cm depth in a water phantom, a 10 cm × 10 cm field size at 100 cm source-to-phantom surface distance. The f_c values were determined as the ratio of absorbed dose to water scored in a fully water-equivalent (dummy) geometry to that scored in a semi-realistic geometry retaining the actual electrode materials - aluminium for the central electrode and graphite-coated PMMA for the outer electrode. As results, the f_c values exceeded unity for all four chambers, ranging from 1.0003 to 1.0138, with combined statistical standard uncertainties ranging from 0.43% to 0.96% (k = 1). The largest correction of 1.4% was observed for the smallest-volume chamber (PTW-31021), demonstrating an inverse dependence between the magnitude of the electrode perturbation correction and the sensitive volume of the chamber. The f_c values are small but non-negligible, particularly for sub-0.1 cm³ ICs. The computed f_c values provide benchmark data applicable to both clinical and metrological dosimetry, confirming the importance of accounting for electrode material perturbations in high-accuracy absorbed dose to water measurements using ICs in clinical photon beams.

版本历史

[V1] 2026-11-18 14:34:02 ChinaXiv:202611.00226V1 下载全文
点击下载全文
预览
同行评议状态
待评议
许可声明
metrics指标
  •  点击量10
  •  下载量1
评论
分享
申请专家评阅
  • 运营单位: 中国科学院文献情报中心
  • 邮箱: eprint@mail.las.ac.cn
  • 地址:北京中关村北四环西路33号
招募志愿者 许可声明 法律声明

京ICP备05002861号-25 | 京公网安备11010802041489号
版权所有© 2016-2026 中国科学院文献情报中心