Abstract:
Accurate determination of the full-energy peak efficiency (FEPE) of high-purity germanium (HPGe) detectors is essential for quantitative gamma-ray spectrometry and numerous applications in nuclear science and technology. In this study, the dependence of the effective photopeak attenuation coefficient used in the Direct Mathematical Method (DMM) on gammaray energy and source-detector distance was systematically investigated for a coaxial HPGe detector. Effective attenuation coefficients were extracted from experimentally determined FEPE values over the energy range 53.16-1770.23 keV at source-detector distances of 5. 877, 10.777, 15.677, 20.677, and 25.677 cm. The extracted coefficients exhibited a pronounced maximum near 122.06 keV, followed by a gradual decrease with increasing energy. In addition, the coefficient magnitude decreased systematically with increasing source-detector distance, with the most significant geometrical influence observed at low energies.To model this behavior, a piecewise logarithmic parameterization was developed, consisting of a second-order logarithmic polynomial for the low-energy region (53.16-122.06 keV) and a fourth-order logarithmic polynomial for the high-energy region (122.06-1770.23 keV). The proposed model reproduced the extracted attenuation coefficients with excellent accuracy, yielding coefficients of determination (R²) between 0.999646 and 0.999843 and root-meansquare errors below 4.6 × 10⁻³ cm⁻¹.Comparison with the total attenuation coefficient of germanium revealed substantial differences in both magnitude and energy dependence, demonstrating that the effective photopeak attenuation coefficient is not an intrinsic material property but rather a detectorresponse parameter that incorporates the combined effects of photon attenuation, detector geometry, and full-energy deposition processes. The fitted coefficients were subsequently parameterized as explicit functions of source-detector distance, leading to a generalized 1