Abstract:
Temperature variations can significantly affect the spectrometric performance of NaI(Tl) scintillation detectors, causing photopeak displacement and compromising energy calibration stability. This study characterises the temperature-dependent spectral response of two nominally identical 2-inch NaI(Tl) scintillation detectors and evaluates the influence of thermal insulation on their performance.Measurements were conducted using Cs-137 and Co-60 gamma-ray sources over an overall temperature range of approximately 12-43 °C. One detector was characterised under non-insulated conditions, while the second was evaluated in both non-insulated and thermally insulated configurations. The analysis focused on the variation of photopeak centroid positions as a function of temperature. Additionally, a separate Co-60 heatingcooling experiment was performed to assess the reversibility of the temperature-induced spectral changes.The results revealed a systematic shift of photopeak centroids towards lower multichannel analyser (MCA) channels with increasing temperature for all three gamma-ray emissions investigated. For Detector 2 without insulation, the absolute centroid displacement over the 16-40 °C interval was greater for the Co-60 photopeaks than for the Cs-137 photopeak. Although both detector units exhibited the same overall temperaturedependent trend, differences in their individual responses were observed, highlighting the need for detector-specific characterisation. The comparison of insulated and noninsulated configurations of Detector 2 yielded similar centroid-versus-temperature regression slopes, with reductions in the absolute temperature coefficients ranging from approximately 3.1% to 4.3%. These results indicate that thermal insulation provides only a modest reduction in temperature sensitivity and does not eliminate the temperature dependence of the spectrometric response under the investigated conditions. Furthermore, the heating-cooling experiment showed that the initial photopeak centroid positions were not fully recovered after the detector returned to its starting temperature, suggesting a possible dependence on thermal history.Overall, these findings demonstrate the importance of temperature monitoring and individual detector characterisation for maintaining the energy stability of NaI(Tl) gamma-ray spectrometry systems. Thermal insulation may provide a complementary means of managing thermal effects, but should not be regarded as a substitute for appropriate temperature-dependent calibration or spectral correction procedures.